007-5

(5% phenyl)
methylpolysiloxane

  Bonded and Crosslinked
  Excellent thermal stability
  Non polar
  Comparable to: SE-52, SE-54, OV-73
  Similar phases: DB-5, Rtx -5, HP-5, Ultra-5, SPB-5, CPSil-8CB, AT-5, BP-5 
 

The 007-5 phase is a (5% phenyl) methylpolysiloxane polymer.  Like the 007-1dimethylpolysiloxane phase, the  007-5 phase is a widely used general purpose GC phase ideal for a broad range of applications.  The presence of the phenyl groups provides induced dipole interactions leading to degrees of increased retention for aromatic solutes.  A 5% phenyl substitution causes a nominal increase in polarity;  yet the 007-5 phase displays the attributes of a non-polar phase.  

The non-polar 007-5 phase, which separates compounds according to boiling point, is another frequently used phase type in GC. Bonding and crosslinking this general purpose GC phase increases the resistance to degradation by ri nsing, large solvent injections and the deposition of non-volatiles.  The 007-5 phase offers excellent efficiency and thermal stability.  In addition, this non-polar phase is less susceptible to oxidation and hydrolysis than phases incorporating more polar functional groups.  

007-5 PHATTM Phase Columns
Similar to the 007-1 dimethylpolysiloxane phase, the 007-5 is offered in very thick films as noted below.  In addition to the standard range of column configurations offered with the 007-5 phase type, a new coating technology developed at Quadrex allows us to extend the functional range of columns coated with this phase type by the application of very thick films of the 007-5 phase.  For example, films up to 3.5 microns on 0.10mm I.D. fused silica and 18.0 microns on 0.53mm I.D. fused silica are now possible.  These PHAT PhaseTM thick film columns are ideal for low molecular weight analyses previously performed on packed columns. 

Advantages of PHAT PHASETM fused silica capillary columns include:
greater sample size for trace analyses of very volatile compounds
increased retention of volatile compounds leading to higher elution temperatures

GC analyses of gases and other types of low molecular weight compounds have been traditionally accomplished on packed columns with heavy loadings or adsorbent type columns.  More recently, porous layer open tubular (PLOT) capillary columns have been utilized in the separation of low molecular weight applications despite the drawbacks in using these types of columns such as sample adsorption and bead migration within the column.  

Attempts have been made to use 'thick filmed' wall coated open tubular (WCOT) capillary columns but the films have typically been confined to a maximum of 8.0 microns due to limitations in conventional column coating technology.  While these 'traditional' thick film columns have been used for some for low molecular weight applications, they could not attain the separations and sample loadings that could be achieved on packed columns.  

By employing a newly developed proprietary coating technique which we call Polymer Height Augmentation Technology (PHATTM), we are now able to reproducibly manufacture fused silica capillary columns having phase ratios comparable to heavily loaded packed columns.  The PHAT PhaseTM columns are available with 007-1, dimethylpolysiloxane and our 007-5, 5% phenyl methylpolysiloxane phases in lengths up to 30 meters.  These PHAT PhaseTM thick film capillary columns are ideal for low molecular weight analyses previously performed on packed columns.  

Application areas for these PHAT PhaseTM columns include the analysis of natural gas, auto emissions, light hydrocarbon refinery cuts, amines, alcohols, free fatty acids, refrigerants, residual solvents and air pollutants as defined in the EPA TOC Methods. 

AVAILABLE COLUMN CONFIGURATIONS
I.D. (mm) Lengths (Meters) Film thicknesses (microns)
0.10 10, 15, 20, 25 0.1, 0.25, (0.5 on 10M. and 15M. only)
0.18 10, 15, 25, 30, 40, 50 0.1, 0.25, 0.5
0.25 12, 15, 25, 30, 50, 60, 75, 100 0.1, 0.25, 0.33, 0.5, 1.0
0.32 10, 15, 25, 30, 50, 60, 75, 100 0.1, 0.25, 0.33, 0.5, 0.52, 1.0, (2.0, 3.0, 4.0, 5.0 on 60M. or less)
0.53 10, 15, 25, 30, 50, 60, 75, 100 0.1, 0.25, 0.5, 0.88, 1.0, 2.0, 2.65, (3.0, 4.0, 5.0, 6.0, 7.0, 8.0 on 50M. or less)

 007-5 PHAT TM Phase extra thick film columns

0.10 10, 20 1.0, 3.5
0.18 15, 20, 30 1.0, 3.0, 5.0, 6.0
0.25 10, 15, 25, 30 3.0, 5.0, 8.0
0.32 10, 15, 25, 30 8.0, 10.0, 12.0
0.53 10, 15, 25, 30 10.0, 12.0, 15.0, 18.0

 Inquire with tech@quadrexcorp.com about custom lengths and films.

We have now added our complete fused silica and Ultra-ALLOY stainless steel column price lists. The lists are sorted by I.D. then column length. You can
CLICK HERE
to go directly to column pricing, or can link to the pricing pages on the individual column information pages.

Click HERE to go directly to the 007-5 price page.

The columns listed are those that are commonly requested by our customers. If you do not see the column configuration you require, CONTACT US, we will be happy to quote.

OPERATING TEMPERATURE RANGE 

film thickness temperature range
up to 1.0 micron -50° C to 350° C
2.0 to 5.0 micron up to 300° C
6.0 to 8.0 micron up to 280° C
9.0 to 12.0 micron up to 240° C
12.0 and above up to 200° C

007-5 Bonded Phase Retention Index
Phase benzene 1-nitropropane 2-pentanone pyridine butanol

007-5

665.62

751.12

689.59

756.23

662.46

 


Every Quadrex capillary column undergoes stringent testing utilizing either a modified Grob test mix or application-specific test mix prior to shipping. An actual Capillary Column Test Data Sheet, which includes the test chromatogram, operating conditions, and general column information accompanies every column we ship.  For more information on the Grob Test, click here.

Below are the results of a 007-5 column tested with the Grob Test mix which shows the excellent level of deactivation achieved with this phase.

Column: Bonded 5% phenyl methylpolysiloxane
25M. x 0.25mm I.D. x 0.5um film
Cat. No.: 007-5-25-0.5F

Temperature: 40ē (6ēmin.) - 210ē (15ēmin.) - 320ēC
Flow: 25 cm/sec, Helium
Injector: 225ēC
Detector: 325ēC, FID

ELUTION ORDER

1. 2,3-butandiol 7. undecane
2. decane 8. 2,4-dimethylaniline
3. 1-octanol 9. methyl decanoate
4. 2,6-dimethylphenol 10. methyl undecanoate
5. nonanal 11. dicyclohexylamine
6. ethylhexanoic acid 12. methyl dodecanoate

 


Send mail to Webmaster@quadrexcorp.com with questions or comments about this web site.
Copyright Š 2002 Quadrex Corporation
Web Site Designed By Bob White From Duet Productions Ltd.

(We are not affiliated in any way with the Quadrex Corporation (Symbol: QUAD) that appears on the stock market.)