SRI NPD DETECTOR
|
 |
Very selective to
nitrogen and phosphorous |
 |
Detects levels to 10ppb |
 |
 |
Exceptionally rugged NPD bead |
 |
Similar in design to
an FID |
|
|
The Nitrogen Phosphorus Detector ( NPD ) is
similar in design to the FID, except that the
hydrogen flow rate is reduced to about 3 ml/min
and an electrically heated thermionic bead ( NPD
bead ) is positioned near the jet orifice.
Nitrogen or phosphorus containing molecules
exiting the column collide with the hot bead and
undergo a catalytic surface chemistry reaction.
The ions created in this reaction are attracted
to a collector electrode, amplified, and output
to the data system.
The NPD is commonly used to detect pesticides,
herbicides, and drugs of abuse because it
responds to N-P compounds about 100,000 stronger
than normal hydrocarbons, making it very
selective. When combined with the
chlorine/bromine selective DELCD, the combined
detector is ideal for pesticide screening since
the NPD selectively detects the organo-phosphate
pesticides, while the DELCD detects only the
chlorinated species. Together all pesticides can
be detected, and the relative responses can help
to identify pesticide species.
The SRI ceramic NPD bead is exceptionally rugged
and long-lasting, offering service from 100-1000
hours depending on operating conditions.
|
|
#SRI-8690-0015,
NPD DETECTOR |
$3,089.00 |
|
SRI NPD/DELCD
COMBINED DETECTOR
|
 |
Very selective to
nitrogen and phosphorous |
 |
Detects levels to
10ppb |
 |
Exceptionally rugged
NPD bead |
 |
Similar in design to
an FID |
|
|
|
#SRI-8690-2615,
NPD/DELCD COMBINED DETECTOR |
$6,179.00 |
|
|